Cation Sodium Analyzer HK-358
- Applications:
+ Power
+ Chem or Petrochemical
+ Pulp and Paper
+ Semiconductor
+ Cation exchange breakthrough,for high acid samples,increase the effective run time on cation exchangers,drum boiler water
- Features
+ Sensitivity to ppb range
+ Advanced embedded MCU technology,SMT and integrative design
+ Panel mount,aluminum enclosure anti-interference
+ 192×64 lattice LCD display
+ User interface language:English
+ Simultancous display of multtiparameters
+ Automatic or manual temperature compensation
+ Two analog outputs and two alarm relay outputs
+ CAN bus communication
+ Large datalog of previous measurements and trend curve trackable
+ Universal power supply for worldwide operation(100~240)V
+ Low reagent consumption
+ Fast response,accurate measurement ease-of-use with minimal maintenance
Specifications
TRANSMITTER
- Measurement Range : (0.00 to 9.36)pNa, 0.01μg/L to 23.0g/L, (-1.000 to +1.000)mV
- Resolution (min) : 0.01pNa, 0.01ug/L, 0.1mV
- Accuracy : ±0.01pNa, 1character, ±0.5mV
- Stability : ±0.01pNa/24h, ±1character/24h, ±0.5mV/24h
- Analog Output : (0~10)mA or (0~20)mA or (4~20)mA,two isolated outputs
- Output Accuracy : ±1%F.S
- Max.load Resistance : ≤800W
- Power Supply : (100~240)VAC, (45~65)Hz
- Power Consumption : ≤8W
- Insulation Strength : Power to ground 1500V, no flying arc in 1 min
- Alarm Output
+ Two alarm relays. Each relay can be configured as: high concentration alarm, low concentration alarm, high temperature alarm and low temperature alarm.
+ One relay can be simultaneously selected by two alarms.
+ If anyone of the two alarms occurs, it will be activated.
- Communication : CAN Bus (option)
- Data Logger : Record measurement values, history events and calibration results
INSTRUMENT
- Measurement Range : (1.00-8.00)pNa, 0.23ug/L-2300mg/L
- Resolution (min) : 0.01pNa or 0.01ug/L
- Accuracy
+ ±0.02pNa or ±0.5ug/L ( the larger one),
+ : whichever is greater (between ±10oC of calibration temperature)
- Response Time (T90) : 90%within 2 min
- Temperature Sensor : Pt1000
- Temperature Range : (0.0 to 99.9)oC
- Temperature Accuracy : ±0.5 oC
- Temperature Compensation : (0.1 to 60.0)oC (auto or manual)
- Calibration Method : Sample calibration and standard solution calibration
- Overall Dimension (mm) : 450(L) × 690(W) × 190(D)
- Cutout Dimensions (mm) : 410 × 645
- Weight : 19kg
- Sample Condition
+ Temperature: (5 to 50)oC
+ : Pressure: (50-300)KPa
+ : Flow: (6-40)L/h
+ : Solid contents in sample: ≤5mm (no gel and grease)
- Ambient Temperature : (5 to 45)oC
- Ambient Humidity : ≤90%RH (no condensation)
- Storage Temperature : (-25 to 55)oC (above 0oC for electrode)